Download Catalog For :

Anesthesia Machine MD-AM-1001

Anesthesia Machine MD-AM-1001

Anesthesia Machine MD-AM-1001 includes 5 tube mechanical flowmeters for accurate gas delivery. It supports O₂, N₂O, and AIR, with flow rates ranging from 0.1 to 10 L/min. Equipped with an ACGO, it ensures rapid oxygen supply between 25 L/min to 75 L/min. It features an integrated breathing circuit design. provides apnea ventilation settings, and suffocation warnings, and displays errors on the screen. Our Anesthesia Machine supports multiple working modes, including volume control and pressure limit, catering to diverse patient needs.

Features:

  • Multiple working modes
  • 5 tube mechanical flowmeters 
  • Integrated breathing circuit design
  • High precision flowmeter
  • Anesthesia Circuit Gas Outlet (ACGO) feature
  • 10.4-inch TFT LCD screen display
  • Alarms for power, voltage, volume, and pressure
  • Supports adult and pediatric patients

Applications:

Our Anesthesia Machine delivers anesthesia for both adult and pediatric surgical procedures in various medical settings.

Specifications:

Ventilation Mode IPPV, A/C, SIMV, SIGH, MANUAL
Ventilator Flowmeter O₂ (0.1 to 10 L/min) N₂O (0.1 to 10 L/min) AIR (0.1 to 10 L/min)
Ventilator Rapid Oxygen Supply 25 L/min to 75 L/min
Ventilator Tidal Volume 0.20 mL to 1500 mL
Ventilator Frequency 1/min to 100/min
Ventilator Pressure triggering Sensitivity -20 cm H₂O to 0 cm H₂O (Based on PEEP)
Ventilator PEEP 0 cm H₂O to 30 cm H₂O
Ventilator Flow Trigger Sensitivity 0.5 L/min to 30 L/min
Ventilator SIGH 0 (off) 1/100 to 5/100
Ventilator Apnea Ventilation OFF, 5s to 60s
Ventilator Pressure Limit 20 cm H₂O to 100 cm H₂0
Monitoring Frequency 0 /min to 100 /min
Monitoring Volume 0 mL to 2000 mL
Monitoring MV 0 L/min to 100 L/min
Monitoring Oxygen Concentration 15% to 100%
AC power failure Alarm Power failure or no connection
Low Voltage alarm for backup battery < 11.3 ± 0.3 V
Volume (Tidal) Alarm ≤ 5 mL within 6s
Oxygen Concentration Alarm 19% to 100% (HIGH) 18% to 99 % (LOW)
Airway Pressure Alarm 20 cm H₂O to 100 cm H₂O (HIGH) 0 cm H₂O to 20 cm H₂O (LOW)
Minute Volume Alarm Adult (5 L/min to 20 L/min) Child (1 L/min to 15 L/min) (HIGH); 0 to 10 L/min (LOW
Continuous Pressure Alarm (PEEP + 1.5 kPa) over 16s
Suffocation Warning 5s to 60s no spontaneous ventilation
Maximum Limited Pressure < 12.5 kPa
Fan Error Show On Screen
Oxygen Deficit Show On Screen
Gas Source O₂, N₂O, AIR
Pressure 280 kPa to 600 kPa
Power Supply 100V to 240 V, 50 / 60 Hz
Dimensions (L × W × H) 920 × 970 × 1380 mm
Net Weight 102 Kg
Gross Weight 156 Kg