Anesthesia System MD-ANS-1006 offers precise gas flow rates for O2, N2O, and air. It ranges from 0.2–15 L/min for O2 and air, 0.2–12 L/min for N2O. Advanced Oscillogram Monitoring tracks critical respiratory parameters. The alarms for battery status, low oxygen supply, and fan errors ensure uninterrupted power. It features adjustable tidal volume (10–1500 mL) and frequency (1–100 breaths/min) for customized ventilation strategies. Our system ensures enhanced safety and flexibility ideal for anesthesia and critical care settings.
Features:
Patient Centered Ventilation
customizable ventilator parameters
Controlled and spontaneous breathing support.
apnea and suffocation alerts.
Advanced Oscillogram Monitoring
Features Real-time monitoring
15” TFT LCD touchscreen
Pressure and Flow Controls
Automatic decompression to ensure safety.
LED top light
Battery Backup
Packed in Compact wooden case
Standard accessories:
Anesthetic breathing system × 1
Power cord 4m × 1
Fuse × 8
Fuse × 1
O2 cell × 1
Adult Anesthesia Breathing Circuit(reusable) × 2
Child Anesthesia Breathing Circuit(reusable) × 2
Breathing tube × 1
Y connector × 1
Y connector × 1
L connector × 2
Straight connector × 1
Reservoir bag 1L × 1
Reservoir bag 3L × 1
Anesthetic mask adult × 1
Anesthetic mask infant × 1
Vaporizer × 2
High pressure O2 hose × 1
High-pressure AIR hose × 1
Optional accessories:
Spare Cylinder yoke, 1 O2, includes gauges × 1
Spare Cylinder yoke, 1 N2O, includes gauges × 1
Disposable Adult breathing circuit × 1
Disposable Pediatric breathing circuit × 1
Anesthetic gas scavenging system × 1
Patient monitor × 1
Applications:
Anesthesia system MD-ANS-1006 performs excellently in ICUs, operating rooms, and anesthesiology departments. It is professionally designed for adult, child, and infant inhalation anesthesia and respiratory management, featuring advanced ventilation modes.
Specifications:
Ventilation Modes | IPPV, PCV-PC, PCV-VG, V-A/C, P-A/C, P-SIMV, V-SIMV, PSV, PCV-VG SIGH, MANUAL |
---|---|
Gas source | O2, N2O, AIR |
Flowmeter Range | O2: 0.2 to 15 L/min N2O: 0.2 to 12 L/min AIR: 0.2 to 15 L/min |
Rapid Oxygen Supply | 25 L/min to 75 L/min |
Oxygen Concentration Range | 15 % to 100 % |
Inspiratory Ratio (I) | 4:1 to 1:8 |
Tidal Volume (Vt) range | 0, 10 mL to 1500 mL |
Minute Volume (MV) Range | 0 L/min to 100 L/min |
Frequency Range | 0 /min to 120 /min |
Positive End-Expiratory Pressure (PEEP) | 0 cmH2O to 30 cmH2O |
Pressure Control Range | 5 cmH2O to 70 cmH2O |
Pressure Triggering Sensitivity (PTr) | -20 cmH2O to 0 cmH2O |
Flow Trigger Sensitivity (FTr) | OFF, 0.5 L/min to 30 L/min |
Apnea Ventilation | OFF, 5% to 60% |
SIGH | 0 (off) 1/100 to 5/100 |
Gas Pressure Range | 280 kPa to 600 kPa |
Pressure Limit | 10 cmH2O to 100 cmH2O |
Bellow | 1L |
Oscillogram Types | P-T (Pressure-Time) F-T (Flow-Time), V-T (Volume-Time) P-V loop (Pressure-Volume Loop) ETCO2-T (ETCO2-Time) F-V (Flow-Volume Loop) F-P (Flow-Pressure Loop) |
High-minute volume alarm | Power supply Undervoltage alarm |
Low-minute volume alarm | Low tidal volume alarm |
High airway pressure alarm | Continuous pressure alarm |
Low airway pressure alarm | High-frequency alarm |
No tidal volume | Suffocation alarm |
High tidal volume alarm | Backup power alarm |
Absorber disconnection alarm | Battery exhausted alarm |
High oxygen concentration alarm | Fan error |
Low oxygen concentration alarm | Low oxygen supply alarm |
High ETCO2 concentration alarm | Low oxygen supply alarm |
Power Consumption | 85 VA |
Power Supply | Voltage: 100 to 240 V Power Frequency: 50/60 Hz |
Packing Dimension | 1005 × 960 × 1700 mm |
Net Weight | 126 kg |
Gross Weight | 253 kg |