Download Catalog For :
Anesthesia System MD-ANS-1004 equipped with flowmeters for O2, N2O, and AIR, adjustable from 0.1 to 10 L/min. The Real-time monitoring of respiratory parameters with oscillograms and loops supports precise monitoring. It offers ventilation capabilities tailored for both adult and pediatric patients. It combines technology with a user-friendly interface, ensuring safety and efficiency in critical care settings. Our system offers comprehensive monitoring and ventilation capabilities, ensuring Safety and precision.
Features:
Multiple Ventilation Modes
Equipped with flowmeters
10.4"Colorful TFT LCD
Real-time monitoring
User-Friendly Interface
AC power failure alarm
high/low-pressure limits
built-in protections
Built-in Battery Backup
Compact and Light-weight
Standard accessories:
Anesthetic breathing system × 1
Power cord 4m × 1
Fuse 50V × 2
Fuse × 1
O2cell × 1
Adult Anesthesia Breathing Circuit(reusable) × 2
Child Anesthesia Breathing Circuit(reusable) × 2
Breathing tube × 1
Y connector × 1
Y connector × 1
L connector × 1
Straight connector × 1
Reservoir bag 1L × 1
Reservoir bag 3L × 1
Anesthetic mask adult × 1
Anesthetic mask infant × 1
Vaporizer × 2
High pressure O2 hose 4.472.018,4 m × 1
High-pressure AIR hose 4.472.026,4 m × 1
Optional accessories:
CO2 sensor × 1
Spare Cylinder yoke, 1 O2, includes gauges × 1
Spare Cylinder yoke, 1 N2O, includes gauges × 1
Disposable Adult breathing circuit × 1
Disposable Pediatric breathing circuit × 1
Transfer and receiving system of exhaust gas × 1
Patient monitor × 1
Applications:
Anesthesia System MD-ANS-1004 is a comfortable and classic choice, and is user-friendly, with its outstanding ergonomic design, this machine prioritizes safety, stability, and convenience, ensuring an enhanced user experience.
Specifications:
Ventilation Modes | IPPV, PCV, A/C, P-SIMV, V-SIMV, PSV, PCV-VG SIGH, MANUAL |
---|---|
Gas Source | O2, N2O |
Flowmeter range | O2: 0 to 10 L/min; N2O: 0 to 10 L/min |
Rapid Oxygen Supply Range | 25 L/min to 75 L/min |
Oxygen Concentration | 15 % to 100 % |
Tidal Volume (Vt) range | 0, 20 mL to 1500 mL |
Frequency range | 1 /min to 100 /min |
Inspiratory Ratio (I) | 2:1 to 1:6 |
Inspirational time | 0.2s to 6s |
Minute Volume (MV) | 0 L/min to 100 L/min |
Positive End-Expiratory Pressure (PEEP) | 0 cmH2O to 30 cmH2O |
Pressure Triggering Sensitivity (PTr) | -20 cmH2O to 20 cmH2O |
Flow Trigger Sensitivity (FTr) | 0.5 L/min to 30 L/min |
SIGH | 0 (off) 1/100 to 5/100 |
Apnea Ventilation | OFF, 5 s to 60 s |
Gas Pressure Range | 280 kPa to 600 kPa |
Pressure Limit | 20 cmH2O to 100 cmH2O |
Bellow | 1.5L |
Oscillogram Types | P-T (Pressure-Time), F-T (Flow-Time), V-T (Volume-Time) P-V loop (Pressure-Volume Loop) F-V (Flow-Volume Loop) F-P (Flow-Pressure Loop) ETCO2-T (ETCO2-Time) |
The AC power failure alarm | Power failure or no connection |
Low voltage alarm for backup battery | < 10.2 ± 0.3 V |
No tidal volume | ≤ 5 mL within 6 s |
High oxygen concentration alarm | 19% to 100% |
Low oxygen concentration alarm | 18% to 99% |
High airway pressure alarm | 20 cmH2O to 100 cmH2O |
Low airway pressure alarm | 0 cmH2O to 20 cmH2O |
High-minute volume alarm | Adult (5 L/min to 20 L/min) |
Low-minute volume alarm | Paed (1 L/min to 15 L/min, 0 to 10 L/min) |
Continuous pressure alarm | (PEEP+1.5 kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | ≤ 12.5 kPa |
Fan error | Show on screen |
Oxygen deficit | Show on screen |
Display | 10.4"Colorful TFT LCD |
Power Consumption | 85 VA |
Power Supply | Voltage: 100 to 240 V Power Frequency: 50/60 Hz |
Packing Dimension | 920 × 970 × 1380 mm |
Net Weight | 87.5 kg |
Gross Weight | 155 kg |