CBC Analyzer
Medical Deals CBC Analyzers deliver comprehensive Complete Blood Count (CBC) assessments with exceptional accuracy and speed. Their high-throughput capability enables the processing of multiple samples per hour, significantly optimizing laboratory workflow. They feature automatic clog removal and real-time self-check functions, ensuring consistent performance with minimal maintenance. Our product delivers superior performance, advanced diagnostics, and user-friendly design for precise hematological investigations.
CBC Analyzer MD-CA-1000
-
Principles : Impedance method for WBC, RBC, and PLT counting; Cyanide-free reagent for hemoglobin test
-
Parameters : 20 parameters: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR
-
Histograms : 3 histograms for WBC, RBC, and PLT
-
Sample Volume : Prediluted mode: Less than or equal to 20μL
Whole blood mode: Less than or equal to 9μL
-
Parameter performance : WBC performance:
Linear Range: 0 to 100 Precision: Less than or equal to 3.5% (4.0 to 6.9); Less than or equal to 2.0% (7.0 to 15.0)
Carryover: Less than or equal to 0.5 %
RBC performance
Linear Range: 0 to 0.800
Precision: Less than or equal to 2.0% (3.50 to 6.5); Less than or equal to 2.0% (3.50-6.5)
Carryover: Less than or equal to 0.5 %
HGB (g/L)
Linear Range: 0 to 280
Precision: Less than or equal to 2.0% (100 to 180)
Carryover: Less than or equal to 0.5 %
MCV (fL)
Precision: Less than or equal to 1.0% (70.0 to 110.0)
Carryover: Less than or equal to 0.5 %
PLT (10⁹/L)
Linear Range: 0 to 1000
Precision: Less than or equal to 5.0% (100 to 500)
Carryover: Less than or equal to 1.0%
-
Throughput : 30 samples per hour
-
Display : 8.4-inch TFT Touch Screen
-
Language : Multi-language support (Chinese, English, Spanish, Portuguese, Russian, French, Bahasa Indonesia)
-
Data Storage : Up to 50,000 results (numeric & graphical information)
-
Communication : LAN Port supports HL7 protocol, Bi-directional LIS
-
Interface : 4 USB ports (for external printer, software upgrade, barcode reader, external WiFi, keyboard, mouse), LAN port (1)
-
Printout : Thermal recorder, 50 mm wide paper, various printout formats
-
Power Requirement : Voltage: 100V to 240V ± 10%
Frequency: (50Hz/60Hz) ±1Hz
Power Consumption: Less than or equal to 180VA
-
Environmental Conditions : Temperature: 15°C to 30°C
Humidity: 10% to 90%
Air Pressure: 70kPa to 106kPa
-
Dimensions : 410 × 300 × 400 mm
-
Weight : Less than or equal to 20Kg
CBC Analyzer MD-CA-1001
-
Principles : Impedance method for WBC, RBC, and PLT counting; Cyanide-free reagent for hemoglobin test
-
Parameters : 20 parameters: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR
-
Histograms : 3 histograms for WBC, RBC, and PLT
-
Sample Volume : Prediluted mode: Less than or equal to 20μL
Whole blood mode: Less than or equal to 9μL
-
Parameter performance : WBC performance:
Linear Range: 0 to 100 Precision: Less than or equal to 3.5% (4.0 to 6.9); Less than or equal to 2.0% (7.0 to 15.0)
Carryover: Less than or equal to 0.5 %
RBC performance
Linear Range: 0 to 0.800
Precision: Less than or equal to 2.0% (3.50 to 6.5)
Carryover: Less than or equal to 0.5 %
HGB (g/L)
Linear Range: 0 to 280
Precision: Less than or equal to 2.0% (100 to 180)
Carryover: Less than or equal to 0.5 %
MCV (fL)
Precision: Less than or equal to 1.0% (70.0 to 110.0)
Carryover: Less than or equal to 0.5 %
PLT (10⁹/L)
Linear Range: 0 to 1000
Precision: Less than or equal to 5.0% (100 to 500)
Carryover: Less than or equal to 1.0%
-
Throughput : 40 samples per hour
-
Display : 10.4-inch TFT Touch Screen
-
Language : Multi-language support (Chinese, English, Spanish, Portuguese, Russian, French, Bahasa Indonesia)
-
Data Storage : Up to 100,000 results (numeric & graphical information)
-
Communication : LAN Port supports HL7 protocol, Bi-directional LIS
-
Interface : 4 USB ports (for external printer, software upgrade, barcode reader, external WiFi, keyboard, mouse), LAN port (1)
-
Printout : Thermal recorder, 50 mm wide paper, various printout formats
-
Power Requirement : Voltage: 100V to 240V ± 10%
Frequency: (50Hz/60Hz) ±1Hz
Power Consumption: Less than or equal to 180VA
-
Environmental Conditions : Temperature: 15°C to 30°C
Humidity: 10% to 90%
Air Pressure: 70kPa to 106kPa
-
Dimensions : 410 × 300 × 400 mm
-
Weight : Less than or equal to 20Kg
CBC Analyzer MD-CA-1002
-
Principles : Impedance method for WBC, RBC, and PLT counting; Cyanide-free reagent for hemoglobin test
-
Parameters : 21 parameters: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC
-
Histograms : 3 histograms for WBC, RBC, and PLT
-
Reagent : M-30D DILUENT
M-30CFL LYSE
PROBE CLEANSER
-
Sample Volume : Prediluted mode: Less than or equal to 20μL
Whole blood mode: Less than or equal to 9μL
-
Throughput : 70 samples per hour
-
Display : 10.4-inch TFT Touch Screen
-
Language : Multi-language support (Chinese, English, Spanish, Portuguese, Russian, French, Bahasa Indonesia)
-
Data Storage : Up to 500,000 results (numeric & graphical information)
-
Communication : LAN Port supports HL7 protocol, Bi-directional LIS
-
Interface : 4 USB ports (for external printer, software upgrade, barcode reader, external WiFi, keyboard, mouse), LAN port (1)
-
Printout : Thermal recorder, 50 mm wide paper, various printout formats
-
Power Requirement : Voltage: 100V to 240V ± 10%
Frequency: (50Hz/60Hz) ±1Hz
Power Consumption: Less than or equal to 300VA
-
Environmental Conditions : Temperature: 10°C to 35°C Humidity: 10% to 90% Air Pressure: 70kPa to 106kPa
-
Dimensions : 410 × 300 × 400 mm
-
Weight : Less than or equal to 20Kg
-
Parameter performance : WBC performance:
Linear Range: 0 to 200 Precision: Less than or equal to 2.0% (7.0 to 15.0)
Carryover: Less than or equal to 0.5 %
RBC performance
Linear Range: 0 to 0.800
Precision: Less than or equal to 2.0% (3.50 to 6.5)
Carryover: Less than or equal to 0.5 %
HGB (g/L)
Linear Range: 0 to 280
Precision: Less than or equal to 1.5% (100 to 180)
Carryover: Less than or equal to 0.5 %
PLT (109/L)
Precision: Less than or equal to 4.0% (150 to 500)
Carryover: Less than or equal to 1.0 %