$8,100
Anesthesia system MD-ANS-1002 offers an Inspiratory to Expiratory Ratio Configurable from 2:1 to 1:6. It provides the ability to fine-tune the breathing cycle. It also provides precise gas flow control for O2, N2O, and AIR, ranging from 0.1 to 10 L/min each. The system is designed for ease of operation with intuitive controls and a digital display. That provides real-time monitoring and alarm notifications. Our Anesthesia system is ideal for modern operating rooms, offering advanced control, flexibility, and safety features that cater to the diverse needs of patients.
Ventilation Modes | IPPV, A/C, SIMV, SIGH, MANUAL |
Gas Source | O2, N2O, AIR |
Flowmeter range | O2: 0.1 to 10 L/min; N2O: 0.1 to 10 L/min AIR: 0.1 to 10 L/min |
Rapid Oxygen Supply Range | 25 L/min to 75 L/min |
Oxygen Concentration | 15 % to 100 % |
Tidal Volume (Vt) range | 0, 20 mL to 1500 mL |
Minute Volume (MV) | 0 L/min to 100 L/min |
Maximum Minute Volume | ≥18 L/min |
Frequency range | 1 /min to 100 /min |
Inspiratory Ratio (I) | 2:1 - 1:6 |
Positive End-Expiratory Pressure (PEEP) | 0 cmH2O to 30 cmH2O |
Pressure Triggering Sensitivity (PTr) | -20 cmH2O to 20 cmH2O |
Flow Trigger Sensitivity (FTr) | 0.5 L/min to 30 L/min |
SIGH | 0 (off) 1/100 to 5/100 |
Apnea Ventilation | OFF, 5 s to 60 s |
Gas Pressure Range | 280 kPa to 600 kPa |
Pressure Limit | 20 cmH2O to 100 cmH2O |
Bellow | 1.5L |
Oscillogram Types | P-T (Pressure-Time), F-T (Flow-Time), P-V loop (Pressure-Volume Loop) |
AC Power Failure Alarm | Yes |
Battery Backup Low Voltage Alarm | Yes; threshold < 11.3 ± 0.3 V |
Tidal Volume Alarm | Triggered if no tidal volume is detected |
High/Low Oxygen Concentration Alarm | 19 % to 100 % (High); 18 % to 99 % (Low) |
Airway Pressure Alarms | High: 20 cmH2O to 100 cmH2O; Low: 0 cmH2O to 20 cmH2O |
Minute Volume Alarms | Adult: 5 L/min to 20 L/min; Pediatric: 1 L/min to 15 L/min |
Continuous Pressure Alarm | Over (PEEP + 1.5 kPa) for 16 s |
Suffocation Warning | 5 s to 60 s with no spontaneous ventilation |
Maximum Limited Pressure | ≤12.5 kPa |
Fan Error, Oxygen Deficit, Power Failure | Displayed on screen |
Display | 10.4"Colorful TFT LCD |
Power Consumption | 80VA |
Power Supply | Voltage: 100 to 240 V Power Frequency: 50/60 Hz |
Packing Dimension | 920 × 970 × 1390 mm |
Net Weight | 87.5 kg |
Gross Weight | 155 kg |
10.4"Colorful TFT LCD
Adjustable ventilation rates
Rapid Oxygen Supply
SIGH: prevent atelectasis
Oscillogram Display
AC Power Failure Alarm
Internal Battery Backup
Low Voltage Alarm
Vaporizer Safety
Suffocation Warning
Power Failure or No Connection Alert
Anesthetic breathing system × 1
Power cord 4m × 1
Fuse × 2
Fuse × 1
O2 cell × 1
Breathing tube × 2
Breathing tube × 1
Breathing tube × 2
Y connector Adult × 1
Y connector Child × 1
L connector × 2
Straight connector (φ22 to φ22) × 1
Reservoir bag 1L × 1
Reservoir bag 3L × 1
Anesthetic mask adult × 1
Anesthetic mask infant × 1
Anesthesia Vaporizer × 1
High pressure O2 hose × 1
High-pressure AIR hose × 1
High pressure N2O hose × 1
CO2 sensor × 1
Transfer and receiving system of exhaust gas × 1
Patient monitor × 1
Anesthesia System MD-ANS-1002 offers high performance in ICUs, operating rooms, and anesthesiology ensuring maximum patient safety and ease of use across adult, child, and infant anesthesia and respiratory care.